Model Sensor Voltage RH Display,3598 T C 115 No,3599 IR 115 No. 3584 T C 115 Yes,3585 IR 115 Yes, T C is a thermal conductivity sensor IR is an infra red sensor. All units are 50 60 Hz,MANUAL NUMBER 7003584, 9 25584 8 6 09 Updated RH low alarm ringback time from 30 minutes to 15 ccs. 8 25248 IN 3945 6 9 09 Updated drawer specs sliding and stationary ccs. 7 24900 IN 3923 4 22 09 Updated door interlock switch 3598 200 9 ccs. 25035 IN 3912 Corrected 200 7 9 drawings from 370 to 3598 ccs. 6 25303 IN 3902 2 17 09 UL decom warning pg 1 7 ccs. 5 24975 10 31 08 Updated water specs on pg 1 10 1 11 ccs. Thermo Fisher Scientific Direct Heat CO2 Incubator i. Important Read this instruction manual Failure to read understand and follow the instructions in this manual. may result in damage to the unit injury to operating personnel and poor equipment performance. Caution All internal adjustments and maintenance must be performed by qualified service personnel. Material in this manual is for information purposes only The contents and the product it describes are subject to. change without notice Thermo Fisher Scientific makes no representations or warranties with respect to this. manual In no event shall Thermo be held liable for any damages direct or incidental arising out of or related to. the use of this manual, Warning If the incubator is not used in the manner specified in this operating manual the protection provided. by the equipment design may be impaired,2006 Thermo Fisher Scientific All rights reserved.
ii Direct Heat CO2 Incubator Thermo Fisher Scientific. Important operating and or maintenance instructions Read the accompanying text carefully. Potential electrical hazards Only qualified persons should perform procedures associated with this. Equipment being maintained or serviced must be turned off and locked off to prevent possible injury. Hot surface s present which may cause burns to unprotected skin or to materials which may be. damaged by elevated temperatures, Marking of electrical and electronic equipment which applies to electrical and electronic equipment. falling under the Directive 2002 96 EC WEEE and the equipment that has been put on the market. after 13 August 2005, This product is required to comply with the European Union s Waste Electrical Electronic. Equipment WEEE Directive 2002 96 EC It is marked with the WEEE symbol Thermo has. contracted with one or more recycling disposal companies in each EU Member State European. Country and this product should be disposed of or recycled through them Further information on. Thermo s compliance with this directive the recyclers in your country and information on Thermo. products will be available at www thermo com, Always use the proper protective equipment clothing gloves goggles etc. Always dissipate extreme cold or heat and wear protective clothing. Always follow good hygiene practices, Each individual is responsible for his or her own safety. Thermo Fisher Scientific Direct Heat CO2 Incubator iii. Do You Need Information or Assistance on,Thermo Scientific Products.
If you do please contact us 8 00 a m to 6 00 p m Eastern Time at. 1 740 373 4763 Direct,1 800 438 4851 Toll Free U S and Canada. 1 740 373 4189 FAX, http www thermo com Internet Worldwide Web Home Page. service led marietta thermofisher com Service E Mail Address. Our Sales Support staff can provide information on pricing and give you quotations We can. take your order and provide delivery information on major equipment items or make. arrangements to have your local sales representative contact you Our products are listed on the. Internet and we can be contacted through our Internet home page. Our Service Support staff can supply technical information about proper setup operation or. troubleshooting of your equipment We can fill your needs for spare or replacement parts or. provide you with on site service We can also provide you with a quotation on our Extended. Warranty for your Thermo Scientific products, Whatever Thermo Scientific products you need or use we will be happy to discuss your. applications If you are experiencing technical problems working together we will help you. locate the problem and chances are correct it yourself over the telephone without a service. When more extensive service is necessary we will assist you with direct factory trained. technicians or a qualified service organization for on the spot repair If your service need is. covered by the warranty we will arrange for the unit to be repaired at our expense and to your. satisfaction, Regardless of your needs our professional telephone technicians are available to assist you. Monday through Friday from 8 00 a m to 6 00 p m Eastern Time Please contact us by. telephone or fax If you wish to write our mailing address is. Thermo Scientific,Controlled Environment Equipment.
401 Millcreek Road Box 649,Marietta OH 45750, International customers please contact your local Thermo Scientific distributor. iv Direct Heat CO2 Incubator Thermo Fisher Scientific. Warranty Notes, Information You Should Know Before Requesting Warranty Service. Locate the model and serial numbers A serial tag is located on the unit itself. For equipment service or maintenance or with technical or special application inquiries contact Technical. Services at 1 800 438 4851 or 1 740 373 4763 USA and Canada Outside the USA contact your local. distributor,Repairs NOT Covered Under Warranty, Calibration of control parameters Nominal calibrations are performed at the factory typically 1 C for. temperature 1 for gases and 5 for humidity Our service personnel can provide precise calibrations as. a billable service at your location Calibration after a warranty repair is covered under the warranty. Damage resulting from use of improper quality water chemicals or cleaning agents detrimental to. equipment materials, Service calls for improper installation or operating instructions Corrections to any of the following. are billable services,1 electrical service connection.
2 tubing connections,3 gas regulators,4 gas tanks,5 unit leveling. 6 room ventilation,7 adverse ambient temperature fluctuations. 8 any repair external to the unit, Damage resulting from accident alteration misuse abuse fire flood acts of God or improper. installation, Repairs to parts or systems resulting from unauthorized unit modifications. Any labor costs other than that specified during the parts and labor warranty period which may. include additional warranty on CO2 sensors blower motors water jackets etc. Thermo Fisher Scientific Direct Heat CO2 Incubator v. Table of Contents,Section 1 Installation and Start Up 1 1.
Keypad Operation 1 3,Displays 1 4,Installing the Incubator 1 4. Stacking the Incubators 1 5,Preliminary Cleaning 1 7. Shelf Installation 1 7,Access Port Filter CO2 Sensor Cover Plate 1 8. Air Sample Filter Installation 1 8,HEPA Filter Installation 1 9. Leveling the Unit 1 9,Electrical Power Connection 1 9.
Filling the Humidity Pan 1 10,Connecting the CO2 Gas Supply 1 11. Incubator Start Up 1 12,Section 2 Calibration 2 1,Temperature 2 1. Thermal Conductivity CO2 System 2 2,Infrared CO2 System 2 3. Relative Humidity 2 4, Thermo Fisher Scientific Direct Heat CO2 Incubator vi. Table of Contents,Section 3 Configuration 3 1,Turn Audible Alarm ON OFF 3 1.
New HEPA Filter 3 1,Set REPLACE HEPA filter reminder 3 2. Set Access Code 3 2,Set Low Temp Alarm Limit tracking alarm 3 3. Enable Low Temp Alarm to Trip Contacts 3 3,Set Low CO2 Alarm Limit tracking alarm 3 4. Set High CO2 Alarm Limit tracking alarm 3 4,Enable CO2 Alarms to Trip Contacts 3 5. Set New Zero Number for T C CO2 Sensors 3 5,Set New Span Number for T C CO2 Sensors 3 6.
Setting a Low RH Alarm Limit 3 6,Enable RH Alarms to Trip Contacts 3 7. Enabling Temp RH to be Displayed 3 7,Section 4 Alarms 4 1. Temp Controller Failure TMP CNTR ERR 4 2,Sensor Fault Alarms 4 2. Section 5 Routine Maintenance 5 1,Clean Cabinet Exterior 5 2. Clean Glass Doors 5 3,Clean Humidity Pan 5 3,HEPA Filter Maintenance 5 5.
Replace Power Fuses 5 5,Replace Air Sample Filter 5 6. Electronics Section 5 6,Sterilization Cycle 5 8, vii Direct Heat CO2 Incubator Thermo Fisher Scientific. Table of Contents,Section 6 Factory Options 6 1,Remote Alarms 6 1. Humidity Readout P N 1900091 6 2,Section 7 Specifications 7 1. Section 8 Spare Parts 8 1,Exploded Part Drawings 8 2.
Section 9 Electrical Schematics 9 1,Section 10 Warranty Information 10 1. Thermo Fisher Scientific Direct Heat CO2 Incubator viii. Section 1 Installation and Start Up,Figure 1 1 Components. Chamber Gas Sample Port Used for sampling chamber CO2 content. using a FYRITE or similar instrument,Main Power Switch Mains Disconnect. Control Panel Keypad displays indicators Figure 1 2. Leveling Legs Used to level the unit, Sterilization Cycle Button Switch to initiate sterilization cycle. Note The incubators are stackable Information follows. Thermo Fisher Scientific Direct Heat CO2 Incubator 1 1. Installation and Start Up,Mutes the audible alarm, Alarm Indicator Light pulses on off during an alarm condition.
MODE Select Switch Used to select Run Setpoints Calibration and. Configuration modes,Message Display Shows system status. Mode Select Indicators,RUN Run Menu,SET Set Points Menu. CAL Calibration Menu,CON Configuration Menu,HEAT Indicator Lights when heaters are activated. INJ Indicator Lights during CO2 injection into chamber. T C Display Programmable to display temperature See. Configuration section, RH Display Shows percentage of humidity See Configuration sec. CO2 Display Programmable to display CO2 percentage See. Configuration section, Arrow Keys Scroll for Parameters Increases or decreases number values.
toggles between mode choices,ENTER Stores the selected value into unit memory. Silence Alarm Mode Selector Active Heat Indicator Temperature ENTER. Indicator Switch Mode Select Indicators in C,Switch button. Message Display, Power Switch Active Injection Indicator Scroll Arrow Keys. Humidity in,CO2 level in,Figure 1 2 Control Panel, 1 2 Direct Heat CO2 Incubator Thermo Fisher Scientific. Installation and Start Up, Keypad Operation The Series 8000DH incubator has four basic modes for incubator setup.
Run Setpoints Calibration and Configuration,Run is the default mode for normal operation. Set is used to enter system setpoints for incubator operation. Calibration is used to calibrate various system parameters. Configuration allows for custom setup of various options. Table 1 1 Selections Under Each Mode,RUN SETPOINT CALIBRATION CONFIGURATION. Default Mode Temperature Temp Offset Audible,Overtemp CO2 Cal1 New HEPA Timer. CO2 IR Cal2 Replace HEPA Reminder,RH Cal Access Code. Temp Lo Alarm,Temp Relay,CO2 Lo Alarm,CO2 Hi Alarm.
RH Lo Alarm,Display Temp,Display RH,T C units Base Unit Displays. IR units Option Unit Displays,T C units only, Thermo Fisher Scientific Direct Heat CO2 Incubator 1 3. Installation and Start Up, Keypad Operation SCROLL FOR PARAMETERS Arrows Steps the operator through. parameters of SET CAL and CON modes The right arrow goes to the. continued next parameter the left arrow returns to the previous parameter. Up Down Arrows Increases decreases or toggles the parameter value. selected in SET CAL and CON modes, ENTER Must press this key to save to memory all changed values. Key Press to mute audible alarm See Section 4 for alarm ringback. Displays Message Display Shows system status Mode at all times Shows CLASS. 100 or SYSTEM OK during normal operation or alarm messages if. the system detects an alarm condition see Section 4 Alarms The mes. sage CLASS 100 is a timing mechanism indicating that under normal. operating conditions with the HEPA filter installed the air inside the. chamber meets the Class 100 air cleanliness standard for particulates of. 0 5 micron size or larger per cubic foot of air Further information on. the Class 100 classification of air quality is available from Thermo. Upper and Lower Displays The 7 segment upper display shows temp and. RH The lower display shows CO2 and O2, Installing the 1 Maintain a minimum three inch clearance behind the incubator for.
Incubator electrical and gas hook ups In addition a three inch ventilation space. is needed on each side, 2 Locate the unit on a firm level surface capable of supporting the unit. weight of 260 lbs 118kg, 2 Locate the unit away from doors and windows and heating and air. conditioning ducts, 3 Lift the unit by the sides of the cabinet base Do not attempt to lift it. by the front and back This places stress on the outer door hinges. 1 4 Direct Heat CO2 Incubator Thermo Fisher Scientific. ii Direct Heat CO2 Incubator Thermo Fisher Scientific Preface Important Read this instruction manual Failure to read understand and follow the instructions in this